Assessment of Velocity Fields and Shear Stresses Distal to Prosthetic Heart Valves Using
High-Field Magnetic Resonance Imaging A high-field magnetic resonance scanner equipped with short echo time phase-contrast velocity measurement software was applied in a detailed mapping of the axial velocity profile across the entire vessel area at two positions downstream of a bileaflet valve prosthesis inserted in a pulsatile flow system in vitro. The laminar shear forces were calculated from the fluid velocity field data. The velocity profiles very close to the valve reflected the bileaflet design as also shown in previous studies, but the extent and velocities of the jet emanating from the slit between the leaflets were clearly better visualized. However, one diameter downstream of the valve the central jet was completely dispersed and the hemodynamic complexity was significantly reduced. Recirculation and retrograde flow regions that might be relevant for understanding typical long-term complications after implantation were observed close to the valve. In one scan experiment the presented method provides information on flow characteristics that previously required application of different types of experiment. |