Share this page on LinkedIn
Share This Page on Google+
Share This Page on Twitter
tell someone about this page print this page
You are here: Contents > 2004 > Volume 13 Number 5 September 2004 > LABORATORY STUDIES > 3-D Simulation of the St. Jude Medical Bileaflet Valve Opening Process: Fluid-Structure Interaction Study and Experimental Validation

3-D Simulation of the St. Jude Medical Bileaflet Valve Opening Process: Fluid-Structure Interaction Study and Experimental Validation

Alberto Redaelli, Helene Bothorel, Emiliano Votta, Monica Soncini, Umberto Morbiducci, Costantino Del Gaudio, Antonio Balducci, Mauro Grigioni
Department of Bioengineering, Politecnico of Milan, Milan, Laboratory of Biomedical Engineering, Istituto Superiore di Sanità, Rome, Italy

Background and aim of the study: Simulation of the opening and closure dynamics of a mechanical valve through a moving deforming mesh algorithm presents a challenge because of the large rotations of the leaflet and of the small gaps between the housing and the leaflets, which make remeshing a critical issue. The present study offers a computational approach to the simulation of valve leaflet motion during the opening process, together with an experimental set-up for validation of the model.
Methods: A fully 3-D simulation of the 27 mm St. Jude Medical Hemodynamic Plus mechanical valve was performed using the computational code, Fluent. Interaction between the leaflet and fluid was simulated through customized user subroutines which, according to a weakly coupled approach, update the leaflet velocities through subsequent time steps by means of an under-relaxation procedure. A parallel, experimental test was defined to collect data for the set-up of simulations and for

validation purposes.
Results: The computed leaflet velocity and angular displacement compared well with experimental data. The model captured the main features of the opening process, and did so also from a quantitative viewpoint. Nonetheless, some discrepancies were observed, including a delay of ~7 ms in the computed leaflet displacement and an underestimation by ~7% of the maximum computed leaflet velocity.
Conclusion: The weakly coupled approach adopted here limited computational costs, thus allowing the simulation of a fully 3-D realistic mechanical valve within 154 CPU hours at minimal computational costs. No significant drawbacks were raised in comparison with the fully coupled approach. The opening process delay was similar to that reported previously, and cannot be ascribed to the weakly coupled approach adopted here.
The Journal of Heart Valve Disease 2004;13:804-813

3-D Simulation of the St. Jude Medical Bileaflet Valve Opening Process: Fluid-Structure Interaction Study and Experimental Validation

Click the above hyperlink to view the article, right click (Ctrl click on a Mac) to open in a new browser window or tab.

Purchase this Article

Please click the button below to purchase this article. Single article purchases are provided at $50.00 per article. Upon clicking the button below, single article user account subscription details are requested and, upon successful payment, a single article user account is created. Single articles are availble in your account for seven days after purchase.