Contents
Download PDF
pdf Download XML
176824 Views
11098 Downloads
Share this article
Review Article | Volume 27 Issue 1 (, 2022) | Pages 87 - 96
Early Hemodynamic Performance of the Crown PRT Aortic Prosthesis: A Prospective Study
 ,
 ,
 ,
 ,
 ,
 ,
 ,
1
Wessex Cardiothoracic Centre, Southampton General Hospital, Tremona Road, Southampton, UK. Electronic correspondence: drsimonduggan@gmail.com.
2
Wessex Cardiothoracic Centre, Southampton General Hospital, Tremona Road, Southampton, UK.
Under a Creative Commons license
PMID : -30560604
Published
Jan. 10, 2018
Abstract

Background: Currently, only limited data are available on the rate of hemodynamic progression with clinical outcome in patients receiving the latest Crown PRT aortic prosthesis. The study aim was to report clinical and hemodynamic outcomes in 55 consecutive patients for a follow up of up to one year after Crown PRT implantation.

Methods: Between February and September 2015, a total of 55 patients (34 males, 21 females; mean age 77.3 ± 1.2 years) underwent aortic valve replacement (AVR) with the latest LivaNova Crown PRT bioprosthesis at the authors' institution. Left ventricular function was preserved in 79% of patients. Data relating to the patients' clinical, echocardiographic and functional capacities were obtained prospectively.

Results: There were no in-hospital deaths. Significant perioperative complications included stroke (3.6%), atrial fibrillation (27%), and permanent pacemaker insertion (1.8%). Pre-discharge echocardiography demonstrated peak (PG) and mean (MG) transprosthetic gradients of 24.4 ± 10.4 mmHg and 12.9 ± 6.2 mmHg, respectively. The Doppler velocity index (DVI) was 0.49 ± 0.13, and the effective orifice area index (EOAi) 0.89 ± 0.12 cm2/m2. At a mean follow up of 1.3 ± 0.3 years, the transprosthetic gradients, DVI and EOAi were not significantly different from postoperative or pre-discharge values. The patients' NYHA status was I or II in 95% of cases, and the mean left ventricular mass had decreased by 36% at the end of follow up.

Conclusions: The Crown PRT is an effective bioprosthesis, with a low incidence of valve-related complications comparable to those of other current bioprostheses. The bioprosthesis demonstrated satisfactory results in terms of hemodynamics and freedom from reoperation.

 

 

 

How to cite: Duggan, S. M., Sadeque, S., Karamanou, D., Salhiyyah, K., Giritharan, S., Pousios, D., Miskolczi, S., & Barlow, C. W. (2018). Early Hemodynamic Performance of the Crown PRT Aortic Prosthesis: A Prospective Study. The Journal of heart valve disease27(1), 87–96.

 
Keywords
Recommended Articles
Research Article
Morphometric Analysis of the Human Mitral Valve: A Cadaveric Study
...
Published: 19/08/2025
Download PDF
Read Article
Research Article
Role of Metformin in Modulating Serum Ghrelin Levels in Obesity
Published: 30/12/2023
Download PDF
Read Article
Research Article
Impact of Metformin Therapy on Serum Adiponectin Levels in Obese Individuals
Published: 14/08/2025
Download PDF
Read Article
Research Article
Prospective Study on Functional Outcome of CTEV By Ponseti Method of Cast Application
...
Published: 31/07/2025
Download PDF
Read Article
© Copyright Journal of Heart Valve Disease