Contents
Download PDF
pdf Download XML
476832 Views
39002 Downloads
Share this article
Case Report | Volume 5 Issue 1 (, 1999) | Pages 117 - 119
Suture loop restriction of the medtronic hall valve: late presentation of a rare complication
 ,
 ,
1
Department of Cardiothoracic Surgery, University Hospital of Wales, Cardiff, United Kingdom.
Under a Creative Commons license
PMID : -8834735
Published
Jan. 22, 1996
Abstract

Restriction of movement of the occluder of a single disc prosthetic valve is a serious and potentially fatal complication. We present a case in which the movement of the disc of a Medtronic Hall aortic valve was restricted by a loop of aortotomy closing suture inadvertently passed around the central guide of the valve.

 

 

 

How to cite: Kuo, J., Rooney, S., & Breckenridge, I. M. (1996). Suture loop restriction of the medtronic hall valve: late presentation of a rare complication. The Journal of heart valve disease5(1), 117–119.

 
Keywords
Recommended Articles
Research Article
A Prospective Study on Heart Rate Variability as an Early Marker of Metabolic Syndrome in Young Adults
Published: 31/12/2024
Download PDF
Read Article
Research Article
Depression and Anxiety as Predictors of Major Adverse Cardiac Events in Patients with Coronary Artery Disease
Published: 30/12/2023
Download PDF
Read Article
Research Article
A Comparative Study on Early Outcomes of Off-Pump and On-Pump Coronary Artery Bypass Grafting (CABG): A Retrospective Study
...
Published: 17/10/2025
Download PDF
Read Article
Research Article
Echocardiographic Evaluation of Pulmonary Artery Hypertension and Left Ventricular Dysfunction in Chronic Obstructive Pulmonary Disease Patients and Its Co-Relation with Severity of Disease
...
Published: 13/10/2025
Download PDF
Read Article
© Copyright Journal of Heart Valve Disease