Comparison of magnetic resonance imaging and Laser Doppler Anemometry velocity measurements downstream of replacement heart valves: implications for in vivo assessment of prosthetic valve function
 ,
 ,
 ,
 ,
 ,
 ,
 ,
1
Cardiovascular Fluid Mechanics Laboratory, Georgia Institute of Technology, Altanta, GA, USA.
Under a Creative Commons license
PMID : -8834728
Published
Jan. 22, 1996
Abstract

Background and aim of the study: The non-invasive, in-vivo assessment of prosthetic valve function is compromised by the lack of accurate measurements of the transvalvular flow fields or hemodynamics by current techniques. Short echo time magnetic resonance imaging (MRI) may provide a method for the non-invasive, in vivo assessment of prosthetic valve function by accurately measuring changes in the transvalvular flow fields associated with normal and dysfunctional prosthetic valves. The objectives of these in vitro experiments were to investigate the potential for using MRI as a tool to measure the complex flow fields distal to replacement heart valves, and to assess the accuracy of MRI velocity measurements by comparison with Laser Doppler Anemometry (LDA), a gold standard.

Methods: The velocity fields downstream of tilting disc, bileaflet, ball and cage, and pericardial tissue valves were measured using both three-component LDA and MRI phase velocity encoding under a steady flow rate of 22.8 l/min, simulating peak systolic flow. The valves were tested under normal and stenotic conditions to assess the MRI capabilities under a wide range of local flow conditions, velocities and turbulence levels. A new short echo time MRI technique (FAcE), which allowed velocity measurements in stenotic jets with high turbulence, was tested.

Results: Good overall agreement was obtained between the MRI velocity measurements and the LDA data. The MRI velocity measurements adequately reproduced the spatial structure of the flow fields. In most cases peak velocities were accurately measured to within 15%.

Conclusions: The results indicate that the FAcE MRI method has the potential to be used as a diagnostic tool to assess prosthetic valve function.

 

 

 

How to cite: Fontaine, A. A., Heinrich, R. S., Walker, P. G., Pedersen, E. M., Scheidegger, M. B., Boesiger, P., Walton, S. P., & Yoganathan, A. P. (1996). Comparison of magnetic resonance imaging and Laser Doppler Anemometry velocity measurements downstream of replacement heart valves: implications for in vivo assessment of prosthetic valve function. The Journal of heart valve disease5(1), 66–73.

 
Keywords
Recommended Articles
Research Article
A Prospective Study on Heart Rate Variability as an Early Marker of Metabolic Syndrome in Young Adults
Published: 31/12/2024
Download PDF
Read Article
Research Article
Depression and Anxiety as Predictors of Major Adverse Cardiac Events in Patients with Coronary Artery Disease
Published: 30/12/2023
Download PDF
Read Article
Research Article
A Comparative Study on Early Outcomes of Off-Pump and On-Pump Coronary Artery Bypass Grafting (CABG): A Retrospective Study
...
Published: 17/10/2025
Download PDF
Read Article
Research Article
Echocardiographic Evaluation of Pulmonary Artery Hypertension and Left Ventricular Dysfunction in Chronic Obstructive Pulmonary Disease Patients and Its Co-Relation with Severity of Disease
...
Published: 13/10/2025
Download PDF
Read Article
© Copyright Journal of Heart Valve Disease