Research Article | Volume 16 Issue 1 (, 2010) | Pages 37 - 41
Silzone-coated St. Jude Medical valves: six-year experience in 46 patients
 ,
 ,
 ,
1
Department of Cardiothoracic Surgery, University Hospital of North Staffordshire, Stoke-on-Trent, United Kingdom. udandekar@yahoo.co.uk
Under a Creative Commons license
PMID : - 17315381
Published
Jan. 30, 2007
Abstract

Background and aim of the study: The Silzone-coated St. Jude Medical valve was developed to reduce prosthetic valve endocarditis, but in the Artificial Valve Endocarditis Trial (AVERT) was recalled following reports of major paravalvular leaks. A separate study suggested an increased risk of thromboembolic complications associated with the Silzone valve. Herein is reviewed the present authors' experience in patients with Silzone valves.

Methods: Between April 1998 and November 1999, 46 patients (28 males, 18 females; mean age 62.6 years; range: 41-78 years) received a total of 49 Silzone valves. Seven of the patients were in NYHA class IV, and 29 in class III; four patients had active endocarditis. Twenty-one patients underwent aortic valve replacement, 22 mitral valve replacement, and three had both mitral and aortic valve replacement. Concomitant procedures were performed in nine patients.

Results: The 30-day mortality was 6.5 % (n = 3). At post-mortem examination, the valves were found to be seated well, with no evidence of malfunction, thrombotic occlusion or infection. Follow up examinations were conducted initially at six months after surgery, and annually thereafter. All patients underwent transthoracic echocardiography. Follow up was 100% complete; the mean follow up period was 73.5 months (range: 62-81 months). Six late deaths have occurred; five deaths were due to non-cardiac causes, and one cardiac death was unrelated to the valve prosthesis. No cases of paravalvular leak have been identified. Two patients had postoperative thromboembolic complications, but echocardiography did not demonstrate the presence of thrombus or paravalvular leakage.

Conclusion: In this single-center, non-randomized study of 46 patients, the findings of increased paravalvular leak with the Silzone valve, as identified in AVERT study, could not be confirmed. Moreover, the incidence of thromboembolic complications reported was modest.

 

 

 

How to cite: Dandekar, U. P., Baghai, M., Kalkat, M., & Ridley, P. D. (2007). Silzone-coated St. Jude Medical valves: six-year experience in 46 patients. The Journal of heart valve disease16(1), 37–41.

Keywords
Recommended Articles
Research Article
Cross Sectional Study on Nutritional Status and Prevalence of Anemia in Rural Adolescents
Published: 26/06/2013
Download PDF
Read Article
Research Article
To Study the Serum Homocysteine Profile Pattern in AMI & Ischaemic Stroke in Young Adults.
...
Published: 30/08/2025
Download PDF
Read Article
Research Article
Incidence, Clinical Profile, Complications and Outcome of Rodenticide Poisoning in Tertiary Care Hospital
...
Published: 26/08/2025
Download PDF
Read Article
Research Article
To Find Out of the Effects of Mother Nutrition Status on Birth Weight, Anthropometric Measurements of Full-Term Newborn
Published: 07/11/2022
Download PDF
Read Article
© Copyright Journal of Heart Valve Disease