Contents
Download PDF
pdf Download XML
299023 Views
20978 Downloads
Share this article
Review Article | Volume 9 Issue 1 (, 2003) | Pages 104 - 111
Different modes of degeneration in autologous and heterologous heart valve prostheses
 ,
 ,
 ,
 ,
 ,
 ,
1
Department of Cardio-Thoracic Surgery, University of Vienna, Austria.
Under a Creative Commons license
PMID : -10678382
Published
Jan. 11, 2000
Abstract

Background and aim of the study: This study was performed to elucidate the mechanism of primary tissue failure of bioprosthetic heart valves, which were fabricated from autologous pericardium (Autogenics). Results were compared with the degeneration pattern of heterologous pericardial bioprostheses.

Methods: Between March 1994 and December 1996, 87 Autogenics heart valves were implanted in the aortic position. Since then, 15 valves had to be explanted due to structural deterioration. The average implant period was 33+/-8 months. All explants were examined by gross morphological evaluation and X-ray analysis to identify the failure mode of these devices. In eight explanted autologous tissue valves and six explanted heterologous pericardial bioprostheses, exact morphological evaluation was performed by scanning electron microscopy, microscopic and immunohistochemical techniques.

Results: All autologous tissue valves failed due to cuspal tears localized at the commissures. Nocalcification could be detected by X-ray analysis and microscopic methods. Endothelial cell coverage was evident at the outflow surface of all autologous bioprostheses. Histological examination showed severe disintegration of the collagen fibers by insudated plasma proteins and erythrocytes, and the absence of the original fibroblasts. Collagen fibers were vigorously altered between the inner and outer stent of the Autogenics valve. In contrast, heterologous pericardial valves failed due to severe calcification of the cusps. Histological evaluation displayed invasion of macrophages and calcific deposits. The collagenous texture of the pericardial tissue was significantly better preserved compared with autologous tissue.

Conclusion: High biocompatibility of autologous tissue valves is indicated by the absence of calcium deposits, macrophages and foreign body giant cells, and the presence of endothelial cell ingrowth. Severe disintegration of autologous tissue suggests that brief immersion in glutaraldehyde generates inadequate mechanical stability of bioprosthetic heart valve material. Heterologous valves exhibit low biocompatibility but superior preservation of the collagenous biomaterial.

 

 

 

How to cite: Grabenwöger, M., Fitzal, F., Gross, C., Hutschala, D., Böck, P., Brucke, P., & Wolner, E. (2000). Different modes of degeneration in autologous and heterologous heart valve prostheses. The Journal of heart valve disease9(1), 104–111.

 
Keywords
Recommended Articles
Research Article
Cross Sectional Study on Nutritional Status and Prevalence of Anemia in Rural Adolescents
Published: 26/06/2013
Download PDF
Read Article
Research Article
To Study the Serum Homocysteine Profile Pattern in AMI & Ischaemic Stroke in Young Adults.
...
Published: 30/08/2025
Download PDF
Read Article
Research Article
Incidence, Clinical Profile, Complications and Outcome of Rodenticide Poisoning in Tertiary Care Hospital
...
Published: 26/08/2025
Download PDF
Read Article
Research Article
To Find Out of the Effects of Mother Nutrition Status on Birth Weight, Anthropometric Measurements of Full-Term Newborn
Published: 07/11/2022
Download PDF
Read Article
© Copyright Journal of Heart Valve Disease