Evaluation of an In-Vitro Thrombosis Assessment Procedure by Application to the Medtronic Parallel™ and St. Jude Medical Valves

Aimee J. Martin, John R. E. Christy
School of Engineering and Electronics, University of Edinburgh, Edinburgh, UK

Background and aim of the study: With the failure of animal trials to detect thrombosis in the Medtronic Parallel valve, interest has returned to finding a suitable in-vitro method for preclinical valve assessment. A technique has been developed that uses renneted milk to detect clotting around heart valves. The study aim was to determine whether the milk test could discern differences in clotting between the Parallel valve and the clinically successful St. Jude Medical (SJM) valve.
Methods: The Parallel valve and the SJM valve were tested in the aortic position of a rigid model heart chamber at 2 l/min, 70 bpm pulsatile flow. Initially, the valves were run for 30 min to obtain data for localized clotting. Subsequently, the valves were run for only 5 min to obtain data for the initial stages of clotting, including clot origin.
Results: Both valves clotted around their hinge regions in the

30-min tests, and this compared well with documented thrombogenic sites in vivo. The extent of clotting was similar, but clot adherence to the hinge pockets was greater on the Parallel valve than on the SJM valve. After the 5-min runs, no clot was visible on the SJM valve, but clot was consistently found in the hinge recesses of the Parallel valve, adjacent to the pivot.
Conclusion: A renneted milk test can discern differences between the clotting potentials of different heart valves. The Parallel valve clotted earlier than the SJM valve, indicating that it was more thrombogenic. Early-stage clotting was seen to occur around the pivot in the hinge pockets. The milk test has potential for contributing to in-vitro preclinical assessment techniques.


The Journal of Heart Valve Disease 2004;13:667-675
Evaluation of an In-Vitro Thrombosis Assessment Procedure by Application to the Medtronic Parallel™ and St. Jude Medical Valves

Click the above hyperlink to view the article, right click (Ctrl click on a Mac) to open in a new browser window or tab.

Purchase this Article

Please click the button below to purchase this article. Single article purchases are provided at $50.00 per article. Upon clicking the button below, single article user account subscription details are requested and, upon successful payment, a single article user account is created. Single articles are availble in your account for seven days after purchase.